Article
Original Article
Hemanth M1, Patil GS2, Karthik J Kabbur*,3, Harieesha KB4, Praveen GR5, Pruthviraj HV6,

1Professor Head,Department of Orthodontics & Dentofacial Orthopedics Dayananda Sagar College of Dental Sciences, Bangalore-560078

2Professor & Ex. HOD,Department of Orthodontics & Dentofacial Orthopedics Dayananda Sagar College of Dental Sciences, Bangalore-560078

3Dr. Karthik J Kabbur, MDS Reader, Department of Orthodontics & Dentofacial Orthopedics, Dayananda Sagar College of Dental Sciences, Bangalore-560078. Mobile – 99000-93738,

4Associate Professor,Department of Orthodontics & Dentofacial Orthopedics Dayananda Sagar College of Dental Sciences, Bangalore-560078

5Assistant Professor,Department of Orthodontics & Dentofacial Orthopedics Dayananda Sagar College of Dental Sciences, Bangalore-560078

6Senior lecturer, Department of Orthodontics & Dentofacial Orthopedics Dayananda Sagar College of Dental Sciences, Bangalore-560078

*Corresponding Author:

Dr. Karthik J Kabbur, MDS Reader, Department of Orthodontics & Dentofacial Orthopedics, Dayananda Sagar College of Dental Sciences, Bangalore-560078. Mobile – 99000-93738,, Email: karthikkabbur@yahoo.com
Received Date: 2016-05-10,
Accepted Date: 2016-06-15,
Published Date: 2016-07-31
Year: 2016, Volume: 8, Issue: 2, Page no. 18-22, DOI: --
Views: 302, Downloads: 4
Licensing Information:
CC BY NC 4.0 ICON
This work is licensed under a Creative Commons Attribution-NonCommercial 4.0.
Abstract

Orthodontic micro-implants have been used as stationary anchorage in recent years. A 3- Dimensional mathematical model was constructed that uses the finite element method, which simulated an orthodontic implant, maxillary 1st molar, maxillary 2nd premolar with its periodontal ligament and cortical and cancellous bone. The stress and displacement on the implant-bone interface were analyzed on different angulations of the implant to the long axis of the maxillary molar (30º to 90º) with a horizontal orthodontic force of 200gms.

Results: As the angulations decreased the von-mises stress and displacement at the cervix of the implant were decreased, at 90º angulations the stress and displacement are 15.91mpa and -9.44 × 10 -4 respectively, and at 300 angulations the stress and displacement are 7.7mpa and -8.50 × 10 -4 respectively.

Conclusions: On the basis of these results, it is concluded that, micro-implants can be safely loaded with 200gms of mesio-distal force. Maximum stress and displacement was recorded when the implant was placed perpendicular to the long axis of maxillary 1st molar.

<p>Orthodontic micro-implants have been used as stationary anchorage in recent years. A 3- Dimensional mathematical model was constructed that uses the finite element method, which simulated an orthodontic implant, maxillary 1<sup>st </sup>molar, maxillary 2<sup>nd</sup> premolar with its periodontal ligament and cortical and cancellous bone. The stress and displacement on the implant-bone interface were analyzed on different angulations of the implant to the long axis of the maxillary molar (30&ordm; to 90&ordm;) with a horizontal orthodontic force of 200gms.</p> <p><strong>Results: </strong>As the angulations decreased the von-mises stress and displacement at the cervix of the implant were decreased, at 90&ordm; angulations the stress and displacement are 15.91mpa and -9.44 &times; 10 <sup>-4</sup> respectively, and at 30<sup>0</sup> angulations the stress and displacement are 7.7mpa and -8.50 &times; 10 <sup>-4</sup> respectively.</p> <p><strong>Conclusions: </strong>On the basis of these results, it is concluded that, micro-implants can be safely loaded with 200gms of mesio-distal force. Maximum stress and displacement was recorded when the implant was placed perpendicular to the long axis of maxillary 1<sup>st</sup> molar.</p>
Keywords
None
Downloads
  • 1
    FullTextPDF
Article
INTRODUCTION

In orthodontic treatment, anchorage control is essential for success1. If there is an imbalance of force over counter force, unwanted tooth movement occurs. A recent development, stationary anchorage (micro-implants) eliminates one of the uncertainties of orthodontic tooth movement by offering absolute control over potentially undesirable counter movements2. A key factor for the success or failure of a dental implant is the manner in which stresses are transferred to the surrounding bone1. Finite element analysis (FEA) allows us to predict stress distribution in the contact area of the implants with cortical bone and around the apex of the implants in trabecular bone3.

One of the main uses of implants is to retract anteriors in maximum anchorage cases. Implants are usually placed at 30º to 60º angulations to the long axis of the teeth in maxillary arch between maxillary 1st molar and maxillary 2nd premolar for en-mass retraction. This will improve retention while reducing the risk of striking a root4.

The objective of the study was to establish a 3-D finite element model for micro-implant and to analyze the influence of different angulations to the long axis of the teeth (30º to 90º) on the biomechanical characteristics of orthodontic anchorage implant-bone interface.

MATERIALS AND METHODS

A 3-dimensional, orthogonal, Cartesian model was generated by using the NISA-II of 64,000 nodes software system. The model was composed of an orthodontic implant with standard dimensions; the implant simulated the position in between upper first permanent and second premolar of average size, with periodontal ligament (PDL) and cortical and cancellous bone. The model was divided into 14,953 nodes and 34,109 elements: all of them were tetra 4-R type, a solid element with 4 faces and 4 nodes. Each node had 6 degrees of freedom—3 rotations and 3 translations. Homogeneous, isotropic, and linearly elastic behavior was assumed for all the materials. Boundary conditions were located at the floor of the nasal cavity and maxillary sinus and at the lateral surface of the maxillary bone.

The implants were simulated in four different angulations of 300, 450, 600 and 900 to the long axis of the maxillary first molar. A simulated orthodontic force, which was 200gms was loaded mesio-distal to the mathematical model. The stresses and deflections were determined at different mesio-distal and occluso-gingival levels. In addition, specific stress values were evaluated in the different nodes of the root surface, PDL, cortical bone, and implant from the apical zone to the cervical margin. The stress and displacement on the implant – bone interface was analyzed.

RESULTS

The behavior of the evaluated structures under single force was linearly or directly proportional to the applied load. Von Mises stresses in the evaluated loads showed the highest stress on the implant and the cortical bone at the cervical third. The lowest stress appeared at the apical third of the implant.

Von Mises stress values at implant-bone interface are show in Table 1, Graph 1 and Fig 1. Displacement values at implant-bone interface are show in Table 2, Graph 2 and Fig 2. The largest stress and deformation was seen in cortical bone and upper region of trabecular bone. Stress and deformation increased as the angulations of the implant to the long axis of the tooth increased.

DISCUSSION

One of the remodeling theories argues that local mechanical signals stimulate or induce regulating cells to trigger remodeling events5. This implies the importance of studying stress levels from orthodontic forces in the root surface, PDL, and cortical bone and those that occur in the anchor unit like bones, or, as in this case, in an Osseo-integrated implant.6

The aim of the present study was to investigate the deformation of the bone surface around an implant in response to force application in different angulations of implant placement. Abnormally high stress concentration in the supporting tissue can result in pressure necrosis and subsequently in implant failure.

FEM models were used to evaluate the load transfer from the mini-screw to the surrounding bone. The primary component of the load transfer takes place at a single revolution of the mini-screw thread with in the cortex. Under the assumed loading condition, the mini-screw is displaced in a tipping mode, causing tensile stress in the direction of the force7. In general stress levels are higher in the cortical bone than in the trabecular bone. The thickness of the cortical bone determines the overall load transfer from the mini-screw to bone and stiffness of the trabecular bone plays only a minor role.8 The cortical surfaces of the maxilla are thinner and less compact than those of the mandible.3 The finite element method was adapted largely to clinical conditions by selecting parameters such as implant and bone shape, stress, and its angulations. By applying the finite element method, the influence of different angulations to the long axis of the maxillary first molar was observed.9

During tooth movement, nonlinear elastic, plastic, and viscoelastic phenomena can occur. In this article, only linear elastic behaviors of the tooth- periodontal structures are considered.10 Therefore, in the future, additional modelling may be needed along with nonlinear elastic analysis. However, the model does provide quantitative results of the complex 3-dimensional stresses caused by mesiodistal forces during orthodontic treatment. The model allows areas of high stress to be accurately located, and these are considered to be relevant in understanding the long-term behavior of orthodontic tooth movement

CONCLUSION

          Micro-implant can be safely loaded with 200gms of mesio-distal orthodontic force.

          As the angulations of the implant to the long axis of maxillary first molar increased stress and deformation was recorded.

          Maximum stress and displacement was recorded when implant was placed perpendicular to the long axis of maxillary first molar.

Supporting File
References
  1. Van Roekel NB. The use of Branemark system implants for orthodontic anchorage: report of a case. Int J Oral Maxillofac Implants.1989;4: 341–344.
  2. Turley PK, Kean C, Schur J, Stefanac J, Gray J, Hennes J, Poon LC. Orthodontic force application to titanium endosseous implants. Angle Orthod.1988;58:151–162.
  3. Clelland NL, Ismail YH, Zaki HS, Pipko D. Threedimensional finite element stress analysis in and around the screw-vent implant. Int J Oral Maxillofac Implants. 1991;6:391–392
  4. Ziegler P, Ingervall B. Optimizing anterior and canine retraction. Am J Orthod Dentofacial Orthop. 1989;95:95–99.
  5. Middleton J, Jones M, Wilson A. The Role of the periodontal ligament in bone modeling: the initial development of a time dependent finite element model. Am J Orthod Dentofacial Orthop.1996; 109:155–162.
  6. Roberts WE, Smith RK, Ziberman Y, Mozsary PG, Smith RS. Osseous adaptation to continuous loading of rigid endosseous implants. Am J Orthod.1984;86:95–110.
  7. Puente MI, Galba´n L, Cobo JM. Initial stress differences between tipping and torque movements. A three-dimensional finite element analysis. Eur J Orthod.1996;18:329–339.
  8. Barbier L, Sloten JV, Krzeinski G, Schepers E. Finite element analysis of non axial versus axial loading of oral implants in the mandible of the dog. J Oral Rehabil. 1998;25:847–858
  9. Wilson AM, Middleton J. The finite element analysis of stress in the periodontal ligament when subject to vertical orthodontic forces. Br J Orthod.1994;21:161–167.
  10. Tanne K, Sakuda M, Burstone CJ. Threedimensional finite element analysis for stress in the periodontal tissue by orthodontic forces. Am J Orthod Dentofacial Orthop. 1987;92:499–505.